Product summary

Dual-channel The PelviTone device possesses all the features of a continence device, in addition to:Advanced multi-phase programs that are divided into time phases with distinct stimulation parameters;

Increased symmetrical stimulation with a reduced pulse width: 50–330 μS.

Three custom programs and eleven preset continuity programs.

The home compliance (mA and time used) is measured using the lock mode function.

NeuroTrac™ VeriProbe – The vaginal probe for electrostimulation and electromyography.
STIM

Dual channel: circuits that are individually isolated.

Amplitude: 0-90 mA into a 500 Ohm load; the actual mA will be less than the indicated mA due to electrode impedance. The maximum will be limited to 86 mA at a 1000 Ohms load (Electrodes in poor condition), and 65 mA at a 1500 Ohms load.

Type: Constant current, maximum output voltage of 180 Volts +10 / -30 Volts

Waveform: A symmetrical, rectangular biphasic wave with zero DC current.Pulse width selection: 50 μS to 330 μS [2% accuracy].
Pulse rate selection: in the continuous mode, the range is 2 to 100 Hz with a 2% accuracy.The treatment duration is adjustable, ranging from one minute to ninety minutes.

Low Battery Indicator: The battery symbol will flash on and off once every second if the battery drops below 6.9 volts +/- 0.2 volts.

Open Electrode Detection: The output current will be reset to zero if an open circuit is detected at the output of channel A or B.

The ramp-up time ranges from 0.1 to 9.9 seconds.

The unit will not operate if the battery voltage is less than 6.6 (+/- 0.2) volts.