Product summary

LAUDA Ultratemp process thermostats facilitate precise temperature regulation of an external temperature within a range of -5 to 60 °C. The application is guaranteed a precise temperature due to the high temperature stability of ±0.5 K. The heat that is released into the environment is reduced by the water-cooled design. The use of non-flammable liquids such as water or water-glycol results in low operating costs and effortless handling.Features:Process thermostat with an Ethernet interface and an integrated heaterProtection level IP 54 (for outdoor installation)

Additional pump for internal circulation
Integrated water filterClear text menu navigation on a graphic LCD
External Pt100 temperature sensor connection br/>Stainless steel housing

Uses conventional refrigerants (HFCs) in compliance with European legislation to regulate F-gases (EU 573/2024) Digital display of pump pressure

Operation is possible through a web server.

A fully electronic continuous controller with PID action.

Industrial block pump with high pressure and high flow rate.

Operates with non-flammable liquids (water, water/glycol).

Condenser cooling water.



Technical information (as per DIN 12876)Ambient temperature range: -15°C to 50°CTemperature stability: 0.5 ± KThe maximum power of the heater is 50 kW.

The maximum discharge pressure at 50 Hz is 4.8 bar.

Pump Pressure Nominal at 50 Hz: 3.5 barThe maximum flow rate (pressure) at 50 Hz is 250 liters per minute.Nominal pump pressure at 60 Hz is 5.7 bar.

Nominal pump flow rate: 180 L/minIn/Outlet connection thread (inside): 1 1/2"Water Cooling Connection Thread Inside: Rp 1 1/2The maximum volume of filling is 100 liters.
The minimum temperature of the cooling water is 0°C.The maximum temperature of the cooling water is 45°C, and the minimum pressure difference of the cooling water is 2 bar.The maximum pressure of the cooling water is 6 bar.