Gold Health Tech Gold Health Tech
search
Available for inquiry Specification-led inquiry

Medical equipment sourcing

Laboratory Thermostat

Reference: 1152889Sample Management
DomainLaboratory, Process
ConfigurationOn Casters
DisplayDigital
Other CharacteristicsHeating, Cooling
manage_search

Product fit review

Compare the available product information with your application and purchasing requirements.

description

Documentation request

Ask which specifications, manuals, or commercial documents are available for review.

support_agent

Supplier coordination

Discuss quantity, destination, packaging, and timing before confirming suitability.

Product information

Product summary

LAUDA Ultratemp process thermostats facilitate precise temperature regulation of an external temperature within a range of -5 to 60 °C. The application is guaranteed a precise temperature due to the high temperature stability of ±0.5 K. The heat that is released into the environment is reduced by the water-cooled design. The use of non-flammable liquids such as water or water-glycol results in low operating costs and effortless handling.Features:Process thermostat with an Ethernet interface and an integrated heaterProtection level IP 54 (for outdoor installation)

Additional pump for internal circulation
Integrated water filterClear text menu navigation on a graphic LCD
External Pt100 temperature sensor connection br/>Stainless steel housing

Uses conventional refrigerants (HFCs) in compliance with European legislation to regulate F-gases (EU 573/2024) Digital display of pump pressure

Operation is possible through a web server.

A fully electronic continuous controller with PID action.

Industrial block pump with high pressure and high flow rate.

Operates with non-flammable liquids (water, water/glycol).

Condenser cooling water.



Technical data (as per DIN 12876)Ambient temperature range: -15°C to 50°CTemperature stability: 0.5 ± KThe maximum power of the heater is 35 kW.

The maximum discharge pressure at 50 Hz is 3.2 bar.

Nominal pump pressure at 50 Hz is 3 bar.

The maximum flow rate (pressure) at 50 Hz is 250 liters per minute.Nominal pump flow rate: 90 L/min
Thread for inlet/outlet connection (interior): Rp 1 1/2

Inside, the water cooling connection thread is Rp 1 1/2.

Maximum volume of filling: 100 litersThe minimum temperature of the cooling water is 0°C.The maximum temperature of the cooling water is 45°C.The minimum pressure difference for cooling water is 2 bar.The maximum pressure of the cooling water is 6 bar.

The flow rate of cooling water is 82.9 liters per minute.

Continue your equipment review

Related products