Gold Health Tech Gold Health Tech
search
Available for inquiry Specification-led inquiry

Medical equipment sourcing

Anesthesia Ventilator

Reference: 1151172 Pediatrics
ApplicationsAnesthesia
TypeElectronic
Patient TypePediatric, Adult
ConfigurationBenchtop

Associated tags

manage_search

Product fit review

Compare the available product information with your application and purchasing requirements.

description

Documentation request

Ask which specifications, manuals, or commercial documents are available for review.

support_agent

Supplier coordination

Discuss quantity, destination, packaging, and timing before confirming suitability.

Product information

Product summary

The Avante Bonair Anesthesia Ventilator is an electronic time-cycled ventilator that includes integrated patient alarms to enhance patient safety. Flow rate, inspiratory time, and expiratory time are all adjustable in the Bonair. Alarms for low and high pressure are adjustable.input:
output:This Bonair Anesthesia Ventilator can be used on the side of an anesthesia machine or mounted on a dovetail bracket for placement on any flat surface. The low pressure alarm indicates the minimum patient circuit pressure that must be achieved with each breath, while the high pressure alarm terminates the inspiratory flow when the predetermined high pressure is reached.input:
output:Accessories & OptionsUser Guide



Bellows Assembly for Adultsinput:
output:Assembly of Pediatric BellowsBellows the drive hose.Drive Hose for Patient CircuitPressure Sensing Hose 2.5m/22mm M/F Connectorinput:
output:Dovetail Bracket for Anesthesia Machineinput:
output:Patient Circuit Pressure Monitor SpecificationsIn general,Height: 240 inchesinput:
output:Length: 300 inchesinput:
output:Depth: 260 inchesinput:
output:Eight kilograms in weightinput:
output:Inspiratory Time: 0.25 to 3.0 secondsinput:
output:Expiratory Time: 0.5-6.0 secondsinput:
output:flow rate: 0.1-1.0 liters per secondAdjustable Pressure Relief: 20-80 cm H2input:
output:I:E Ratio: A minimum 1:1 ratio derived from the ratio of inspiratory and expiratory timeinput:
output:Tidal Volume: 25 ml to 1600 ml for adult bellows; 25 ml to 300 ml for pediatric bellowsinput:
output:Gasoline: Oxygeninput:
output:Supply Pressure: 45-100 PSI (310-700 KPA)input:
output:Airway Pressure Gauge: -20-100 cm H2O Rangeinput:
output:Low Pressure Alarm with Adjustable Range: 0-30 cm H2input:
output:Adaptable High-pressure alarm: 10-70 cm H2input:
output:Mute Button - LED and 15-second delayinput:
output:Visual and audible alarms for patients: For disconnect/low airway pressure, high airway pressure, high continuous pressure, and sub-ambient pressure.

Continue your equipment review

Related products